TECHNICAL demineralizers are widely used water treatment systems fed with tap water, which retain 96-99% of organic and inorganic pollutants and heavy metals dissolved in water. Fully automatic and maintenance-free devices are equipped with a microprocessor-based control and measurement system that constantly monitors all stages of water treatment. The obtained water meets the requirements of PN-EN ISO 3696: 1999, ASTM, CLSI for tertiary purity waters.
TECHNICAL systems are an excellent source of water for dishwashers, analyzers, autoclaves, climate cabinets, water baths, Ultra demineralizers, and by distributing an appropriate network they can supply water to several laboratory rooms and floors in the building.
General information
Power supply: tap water
Productivity: 5 l / h
Purified water feeding speed: 1-2 l / min
Retention rate> 96-99%
Bacteria <1cfu / ml *
Standard: purified water in the device meets the requirements of PN-EN ISO 3696: 1999 for tertiary water purity.
Application: the obtained water is used to prepare nutrients, buffer solutions, reagents, and is a power supply for other laboratory devices – dishwashers, analyzers, autoclaves, climate cabinets, water baths, polishing demineralizers.
* device equipped with a UV lamp
Technical parameters:
The device works under tap water pressure.
Water purification levels:
cascade filtration on sediment filters,
adsorption on an activated carbon bed,
softening bed (anti-scale),
reverse osmosis membrane module.
System efficiency: approx. 5 dm3 / h (120 dm3 daily).
The system is equipped with a pump to increase the supply pressure with automation.
Retention rate: 96-99%.
Maximum filtrate pressure (treated water) – 3 bar (44 psi).
Automatic and maintenance-free operation of the device.
Water intake point equipped with a demi water spout with a range of min. 2 m.
Possibility to install a tank for storing purified water.
Automatic shutdown of the system when the tank is full or the filtrate valve is closed.
Can be connected to a dishwasher, autoclave, washer, climatic chamber, cleaning system, etc.
Possibility to create a demi water network with several water intake points.
Possibility of independent service (easy replacement of filter modules).
Possibility to install a UV sterilizer.
Possibility of self-assembly of the device.
Power supply: 230V / 50Hz.
Functions securing the operation of the system – interrupting the operation of the pump with the solenoid valve at:
low feed water pressure (no feed water) – low pressure sensor,
tank full – high pressure sensor.
Preview of monitoring messages / alarms.
System monitoring functions (with C automation):
The device is equipped with 24V automatics with a control and measurement system with:
color touchscreen graphic display with the Touch Panel function,
conductivity meter measuring the conductivity and temperature of supply water, after reverse osmosis and ultrapure, in units of uS / cm or MOhm, alternatively,
automatic temperature compensation,
clock displaying date and time,
information about the current operating status of the system,
alarm informing about replacement of prefilters,
alarm informing about the replacement of the membrane module,
information on the degree of retention of the membrane module,
preview of service dates,
graphic and sound alarm signaling,
tank filling level,
the possibility of dispensing water,
built-in RS 232 port for communication with a computer ensuring the possibility of individual adjustment of the service frequency and alarm levels,
built-in USB port for communication with a computer ensuring the possibility of individual adjustment of the service frequency and alarm levels,
possibility of data archiving.
Feed water pressure gauge.
Functions securing the operation of the system:
Pump stoppage when:
low feed water pressure (no feed water) – low pressure sensor,
full tank? high pressure sensor.
Thermal protection of the osmotic module operation, automatic shutdown of the system operation when the supply water temperature is below 40C or above 400C.
Possibility of stopping the system operation in case of any alarm.
System autostart.
Preview of monitoring messages / alarms.
Dimensions (W x D x H): 235 x 425 x 510mm
Feed water parameters:
Conductivity <1200 uS / cm
Pressure> 3.5 bar
Temperature: 4-40 ° C
Hardness <250 mg CaCO3 / dm3
Iron <0.2 mg / dm3
Connections at the installation site:
cold tap water connection 1/2 “or 3/4”
230V socket
drain to the sewage system